Xylazine Test strip guide
Click on the arrows below the images to scroll through and follow the steps for testing your supply for Xylazine or its street name “Tranq”.
A copy of the text within the picture is available below each image!
WARNING: This test does not evaluate drug Safety or Purity, and is NOT for testing cocaine samples, due to cross reactivity with lidocaine, a common adulterant of cocaine. This will cause a false positive result.
Created by SUNAR (2023) based on materials on the BTNX website. For more info, please see https://www.btnx.com/

Demonstration on how to use the Rapid Response Xylazine Test Strip

WARNING: This test does not evaluate drug Safety or Purity, and is NOT for testing cocaine samples, due to cross reactivity with lidocaine, a common adulterant of cocaine. This will cause a false positive result.

NOTE: Humidity and temperature can affect results. DO NOT USE after expiration date printed on the back of the pouch. DO NOT USE the test if the foil pouch is damaged

When you're ready to begin the test, dilute the drug in room temperature water.

Add 5mL ( 1 teaspoon) of room temperature water to a clean container.

Add 5-10mg of drug sample to the container with 5ml of water. *Drug amount is roughly the size of a Q-tip head!

Mix the drug in the room temperature water until it is completely dissolved.

When you are prepared to run the test, open the pouch from the indented tear locations provided for you.

Remove the test strip by holding the strip on the colored end where "XYL" is written. Make sure to use the strip as soon as possible.

Dip the test strip into the sample liquid. KEEP THE STRIP IN THE LIQUID .FOR 10-15 SECONDS

You will want to immerse the strip within the wavy lines. DO NOT DIP PAST THE SOLID MAX LINE.

Remove the strip from the sample.

Place the strip on a non-absorbent flat surface.

Read the test results at 5 minutes. DO NOT READ RESULTS AFTER 10 MINUTES. The back of the pouch and the following picture shows how to read the test!